    CONTEC MS200 NIBP Simulator Blood Pressure Monitor Accuracy Simulation Test

    c8abb8af-de9f-429f-b5de-93b07156e143

    $759.99
    c8abb8af-de9f-429f-b5de-93b07156e143

    Data sheet

    Specific References

    Introduction

    It is a multi - purpose test device for testing NIBP Monitor adopted Oscillometry. Concise and elegant in appearance, easy and convenient to operate, and it can generate the pressure up to 400mmHg (53.3 kPa) by the built-in air pump. With the functions of dynamic BP simulation under several conditions, static pressure calibration, automatic leak test and over-pressure relief test, which can quickly and accurately analyze each performance of NIBP Monitor.


    Functions

    1)Pressure leak test
    2)Over-pressure relief test
    3)BP simulation under standard condition
    4)BP simulation under patient condition
    5)BP simulation under arrhythmia condition
    6)BP simulation under respiratory interference condition
    7)BP simulation of neonate
    8)Wrist BP simulation
    9)User-defined BP simulation
    10)System setting function.
    11)The parameters set can be stored.
    12)ON/OFF function
    13)Dynamic BP simulation


    Performance

    BP simulation range:
    Systolic pressure: 2.7 kPa (20 mmHg) ~ 40.0 kPa (300 mmHg)
    Diastolic pressure: 1.3kPa (10 mmHg) ~ 33.3 kPa (250mmHg)
    Pulse volume: 0.1 cc ~ 2.4 cc
    HR range: 30 bpm ~ 250 bpm; pulse volume > 2.00 cc, maximum HR: 200 bpm
    HR accuracy: ± 1 bpm (except patient condition, arrhythmia and respiratory disturbance)
    Under patient condition:
    Weak pulse, tachycardia, obese, geriatric: ±(preset value 1%+1)bpm
    Moderate exercise: ±(preset value 1.5%+1)bpm
    Strenuous exercise: ±(preset value 3%+1)bpm
    Pressure measurement:
    Range: 0kPa (0 mmHg) ~ 53.3 kPa (400 mmHg)
    Resolution: 0.1 kPa, 1 mmHg, 1 cmH2O, 1 inH2O, 0.1 psi
    Accuracy:
    0 kPa (0 mmHg) ~ 40.0 kPa (300 mmHg), ± (0.5 % of reading + 0.1) kPa or ± (0.5% of reading +1) mmHg
    40.1 kPa (301 mmHg) ~ 53.3 kPa (400mmHg), ± 2% of reading
    Pressure leak test
    Generate a pressure up to 400 mmHg by the built-in air pump. It always displays the peak pressure and the real-time pressure. Leak rate is displayed after 10 seconds.
    Over-pressure relief test
    Generate a pressure up to 400 mmHg by the built-in air pump, it displays peak pressure.
    Pressure source:
    Output range: 6.7 kPa (50 mmHg) ~ 53.3 kPa(400 mmHg)
    Resolution: 0.1 kPa, 1 mmHg, 1 cmH2O, 1 inH2O, 0.1 psi
    Deviation between target pressure and actual pressure: ± 1.3 kPa (± 10 mmHg) in range of 13.3 kPa (100 mmHg) ~ 53.3 kPa (400 mmHg) with a minimum volume of 300 cc
    Airtightness
    Total airtightness: < 0.3 kPa/min (2.25 mmHg/min) under “OFF” state
    Airtightness of built-in cuff: 0.27 kPa/min (2 mmHg/min)
    BP simulation preset value configuration table


    BP simulation BP preset value BP HR Pulse volume
    Standard BP simulation #1 120/80(93) 80bpm 0.68cc
    #2 150/100(116) 0.65cc
    #3 200/150(166) 0.60cc
    #4 255/195(215) 0.55cc
    #5 60/30(40) 0.75cc
    #6 80/50(60) 0.71cc
    #7 100/65(76) 0.69cc
    BP simulation under patient condition
    Healthy heart 120/80(93) 75bpm 0.68cc
    Weak Pulse 110/80(90) 95bpm 0.50cc
    Mild Exercise 140/90(106) 120bpm 1.00cc
    Strenuous Exer 140/90(106) 162bpm 1.40cc
    Obese 120/80(93) 90bpm 0.50cc
    Geriatric 150/110(123) 95bpm 0.40cc
    Tachycardia 120/105(110) 130bpm 0.40cc
    Bradycardia 120/60(80) 45bpm 1.10cc
    BP simulation under arrhythmia condition
    Premature Atrial Contraction1 138/53(81) 80bpm 0.70cc


    Premature Atrial Contraction2 144/64(90) 83bpm
    Premature Ventricular Contraction 118/61(80) 83bpm
    Atrial Fibrillation and PVC 139/72(94) 91bpm


    BP simulation under respiratory interference condition
    spontaneous breathing1 138/65(89) 104bpm 0.60cc
    spontaneous breathing2 149/65(93) 105bpm
    spontaneous breathing3 112/47(68) 86bpm
    controlled breathing 132/44(73) 98bpm
    BP simulation of neonate
    #1 35/15(22) 120bpm 0.11cc
    #2 60/30(40) 0.10cc
    #3 80/50(60)
    #4 100/70(80)
    BP simulation of wrist cuff
    #1 120/80(93) 80bpm 0.50cc
    #2 160/100(120)
    #3 80/55(63)
    User-defined BP simulation[1] - - - -
    Note[1]:Support 9 user-defined BP simulation modes. The following parameters can be defined by users:2.3.1.The value of diastolic pressure should be lower than systolic pressure, pulse volume and HR can not be set to the highest values (e.g., pulse volume ≤ 2.00cc, HR: 250 bpm; pulse volume > 2.00cc, HR: 200bpm).


    Accessories

    One User Manual
    Two fuses
    One power cord


    Physical characteristic

    Display: 4.3" 65 K color LCD display
    Power supply: AC100 ~ 240 V 50/60 Hz, <60 VA
    Environment temperature: + 5 ℃ ~ + 40 ℃
    Relative humidity: < 90 %, non-condensing
    Altitude: ≤ 3000 m
    Atmospheric pressure: 80 kPa ~ 106 kPa
    Dimension: 207 (L) × 220 (W) × 121 (H) mm
    Weight: about 2.1 kg